Infiltec logo

Model DL1 Duct Leakage Tester Specifications.

A precision measurement tool for measurement of low pressure
residential duct leakage, or the leakage of any small enclosure
at pressures from 0.01 "wc (4 Pa) to 1.5 "wc (375 Pa).
Call (540) 943-2776 weekdays 7AM-4PM EDT for more information.

rainbow line

| New | Search | Links | Infiltec | Staff | Fun |
RADON CONTROL | About Radon | Radon FAQ | Pictures | Drawings | Fans | Water | Large Buildings | Radon Catalog |
AIR LEAKAGE | About Blower Doors | FAQ | Ducts | Fire | Large Buildings | Contractors | Blower Door Catalog |
OTHER INSTRUMENTS | Personal Seismometers | Infrasound Monitors |Micro-Manometers: Single Channel Dual Channel |

Click on the thumbnails below for larger images.

Infiltec DL1-DM4 Digital Duct Leakage Tester set up for test,
  low pressure measurement of duct leakage flow and area,
  includes calibrated fan and digital gauges in case, flex duct, tubing and manual
  (110 and 220 vac models available) DL1-DM4 set up for duct leakage test Infiltec DL1-DM4 Digital Duct Leakage Tester with calibration orifice DL1-DM4 set up for calibration check

rainbow line

DL1 FAN SPECIFICATIONS
FEATURE SPECIFICATION COMMENTS
Fan Motor 155 Watts @ 110 vac/60 Hz Powerful, light weight and speed controllable (240 vac/50 hz option)
Flow Calibration Pitot traverse array Single flow sensor does not require multiple "rings"
Maximum flow @ 25 Pa 310 cfm or 527 m3/h (0.146 m3/s) leakage areas up to 200 in2 or 1290 cm2
Maximum Flow at higher Pressures Cfm @ "wc (m3/h @ Pa) 300 @ .75 (510 @ Pa) , 260 @ 1.0 (442 @ 250) , 200 @ 1.25 (340 @ 311), 150 @ 1.5 (225 @ 374) , 50 @ 2.0 (85 @ 500)
Minimum calibrated flow for standard DL1 30 cfm or 51 m3/h (0.014 m3/s) leakage areas down to 5 in2 or 32.2 cm2
Minimum calibrated flow with optional 16C orifice tube 5.32 cfm or 9.05 m3/h (0.0025 m3/s) leakage areas down to 0.89 in2 or 5.7 cm2
Minimum calibrated flow with optional 4C orifice tube 1.33 cfm or 2.26 m3/h (0.00063 m3/s) leakage areas down to 0.22 in2 or 1.4 cm2
Minimum calibrated flow with optional 1C orifice tube 0.33 cfm or 0.57 m3/h (0.00016 m3/s) leakage areas down to 0.055 in2 or 0.35 cm2
Pressure or depressure Calibration unchanged Connect flex hose to other side of fan enclosure
Flex Duct With connectors Six inch diameter, 10 foot length
Speed control location With gauges on case Next to flow and pressure gauges
Fan housing Fiberglass (black) Metal fan blade guards on each end
Case length 29 inches (73.7 cm) Flex duct coupling project 1 inch on each side
Case width 24 inches (38.8 cm)
Case weight 38 lbs (17.3 kg) Handle on case for carrying
Maximum current 1.28 amps @110VAC 0.64 amps @220VAC
Shipping box Weight 42 lbs (19.1 kg) Dimensions: 24.5 x 13 x 24.5 inches (62.2 x 33.0 x 62.2 cm)

rainbow line

DL1 CONTROLS
FEATURE SPECIFICATION COMMENTS
DL1-DM4: Dual Digital Micro-manometer Infiltec DM4 micro-manometer Dual channel, electronic, digital, 3"wc (750 Pa) pressure, 0.001"wc (0.1 Pa) resolution, computes flow and leakage area.
Calibration orifice 20 in2 (129 cm2) Field test the DL1 any time for accuracy
Flow display units Either flow in cfm or m3/h or leakage area in in2 or cm2) Select units by pressing "units" key on DM4 keypad
Pressure gauge units Either Pa or "wc Select units by pressing "units" key on DM4 keypad
Pressure calibration 1% of gauge reading or 1 digit Calibration certificate shows actual errors
Maximum pressure displayed 750 Pa and 3.0 "wc two independent pressure sensors
Flow calibration within 5% DL1 calibration certificate shows actual measured accuracy
Speed control Triac solid-state control On gauge panel
Case Contents Gauges, speed control, fan, manual, power cord, misc tools

rainbow line

DL1 FLEX-DUCT
FEATURE SPECIFICATION COMMENTS
Flex Duct diameter 6 inches (15.2 cm) couplings on each end
Flex Duct length 10 feet (3.05 meters) carrying bag included
Duct connector Flat plate, 8 inch (20.3 cm) Attach to duct by taping connector to opening
Flex material Fabric & wire Durable flex-duct
Shipping box 5? lbs (2.7? kg) Dimensions: 40? x 8? x 5? inches (137 x 27.9 x 12.7 cm)

rainbow line

DL1 OPTIONS
FEATURE SPECIFICATION COMMENTS
240 VAC/50 Hz 155 Watt Includes motor and speed controller, without electrical plugs (country dependent)
Low Flow Orifice Tubes Three tubes with orifice flow sensors Model 16C for flows down to 5.32 cfm or 9.05 m3/h (0.0025 m3/s), or areas down to 0.89 in2 or 5.7 cm2,
Model 4C for flows down to 1.33 cfm or 2.26 m3/h (0.00063 m3/s), or areas down to 0.22 in2 or 1.4 cm2,
Model 1C for flows down to 0.33 cfm or 0.57 m3/h (0.00016 m3/s), or areas down to 0.055 in2 or 0.35 cm2
Smoke Puffer Contains titanium tetrachloride Lasts up to 6 months after activation, comes in resealable tube

rainbow line

| New | Search | Links | Infiltec | Staff | Fun |
RADON CONTROL | About Radon | Radon FAQ | Pictures | Drawings | Fans | Water | Large Buildings | Radon Catalog |
AIR LEAKAGE | About Blower Doors | FAQ | Ducts | Fire | Large Buildings | Contractors | Blower Door Catalog |
OTHER INSTRUMENTS | Personal Seismometers | Infrasound Monitors |Micro-Manometers: Single Channel Dual Channel |

These specifications were updated on September 30, 2015, and are subject to change without notice.

This page copyright © 1996-2015 by Infiltec Inc., all rights reserved. It was last updated on October 07, 2016.

Infiltec small logo
Return to
the Infiltec
Home Page.