DM5 Dual Digital Manometer Specifications
Feature | Specifications |
Sensors: | 2 independent differential pressure transducers with auto-zeroing |
Pressure Resolution: |
Display resolution is 1 Pa from -7.5 kPa to +7.5 kPa or 0.01 "wc from -30.00 "wc to +30.00 "wc. Internal resolution is 1/4 Pa or 0.001 "wc and this is the resolution of the serial output. |
Pressure Range: |
Bi-directional: ±30.00 "wc and ±7.500 kPa. For lower pressures, see the Infiltec DM4 |
Auto-Zeroing: |
Each pressure sensor is connected to a pneumatic switch system that measures and adjusts the zero pressure every 10 seconds. Drifts due to temperature and orientation change are removed by this auto-zeroing. Accuracy is specified as a % of measurement rather than as a % full scale due to auto-zero. |
Keys: | 7 mode selection keys. |
- Device Key: | Select: Manometer or Pitot tube modes. |
- Units Key: |
Dual Manometer: Metric (Pa, Pa), U.S. ("wc, "wc), Mixed ("wc, Pa) Pitot flow: Metric (Pa, m/s), U.S. ("wc, fpm), Mixed (Pa, fpm) |
- Time Avg Key: | Select block average of 1 (default), 2, 4, 8, 16, 32 or 64 seconds. |
- Low Flow Key: |
Dual Manometer: no function Pitot flow: no function |
- Baseline Key: | Resets zero pressure. |
- Hold Key: | Holds reading on display until pressed again. |
- On/Off Key: | Turn power on and off. |
Accuracy: |
Dual Pressure: ±1% of display measurement or ±1 digit, whichever is greater. Pitot Flow: ±1.5% of measurement or better from 2540-21000 fpm (12.9-110 m/s) |
Display Type: | 2x16 high contrast LCD with numbers and characters. |
Operating Temperature: | 32°F to 122°F. (0°C to 50°C) |
Storage Temperature: | -4°F to 160°F. (-20°C to 70°C) |
Maximum Pressure: | ±250 "wc, ±60 kPa or ±9 PSI |
Low Battery Display: |
Current battery voltage is displayed at startup, "lowbat" displayed when voltage is under 7 volts. |
Over Range Indicator: | !!MAX!! displayed when pressure is beyond ±30 "wc or ±7.5 kPa. |
Auto Zero: | Pneumatic switches update zero pressure estimates every 8 seconds. |
Weight With Battery: | ~16 oz. (~455 grams) |
Dimensions: | 4.0"x7.5"x1.79" (101.6x190.5x45.4 mm). |
Battery Life: | Over 100 hours from six AA alkaline batteries. |
Operating Medium: | Air or non-corrosive, non-explosive gas. |
Pressure Connections: | Two pairs, barbed for 1/8" ID tubing. |
Serial Output: |
9600-8N0 pseudo RS232 ASCII records at 1 Hz with LF+CR termination ±32767 range, 1/4 Pa resolution, comma separates two channel values |
Microprocessor: |
Programmable 4 Mhz Microchip PIC14000 with 16 bit a/d. DM5 functions can be updated by replacing this chip. |
Custom Firmware: |
Custom microprocessor programming is available to optimize the DM5 for your pressure/flow/ELA measurement application. |
Optional Accessories: | Pitot tube, serial cable, 110VAC power adaptor. |
Accessories Included: | Battery, soft carrying case, tubing, calibration certificate, user manual. |
Warranty: | 2 years coverage on parts and labor to repair any defects. |
For current DM5 prices, see the Infiltec Blower Door catalog, or call (540) 943-2776 weekdays 7AM-4PM ET.
This page copyright © 1996-2019 by Infiltec Inc., all rights reserved. It was last updated on January 20, 2019.